Underway Chloro and Phaeo Pigments,
Hull Intake Sampling
Methods:
Surface chlorophyll samples were collected from the ship's flow-through seawater system, which has seawater intakes between 5 and 7 meter depths on both ships. About one liter of seawater was vacuum filtered onto GF/F filters and passively extracted in 7 ml of 90% acetone at -20deg C in the dark for at least 24 hours. Chlorophyll fluorescence was then measured on a Turner Design Digital 10-AU-05 fluorometer calibrated prior to the cruise.
Calibration Values
Southern Ocean GLOBEC II - Nathaniel B. Palmer 01-04
Notes:
Performed serial dilution calibration at start of cruise using Sigma Chemicals chl_a standard.
The following are the calibration values:
|
Fd |
|
tau |
H |
0.109 |
H |
1.970 |
M |
0.111 |
M |
1.950 |
L |
0.109 |
L |
1.928 |
Southern Ocean GLOBEC III - Laurence M. Gould 02-03
Notes:
Performed serial dilution calibration at start of cruise using Sigma Chemicals chl_a standard.
The following are the calibration values:
|
Fd |
|
tau |
H |
0.145 |
H |
2.301 |
M |
0.143 |
M |
2.317 |
L |
0.145 |
L |
2.501 |
Study area:
Mean: 1.15
SD:0.49
Median:1.10
Range:0.062-2.40
Southern Ocean GLOBEC IV - Nathaniel B. Palmer 02-04
Turner Designs 10-AU Fluorometer
Notes:
Performed serial dilution calibration at start of cruise using Sigma Chemicals chl_a standard.
The following are the calibration values:
|
Fd |
|
tau |
H |
0.089 |
H |
2.276 |
M |
0.086 |
M |
2.270 |
L |
0.090 |
L |
2.384 |
Data Contributed by: