PI: Ralf Goericke
of: University of California, San Diego
dataset: Pigments, HPLC method, standard casts
dates: February 16, 1998 to March 14, 1998
location: N: -52.9662 S: -71.3072 W: -174.7325 E: -165.9148
project/cruise: AESOPS/RR_KIWI_9, APFZ Polar Front Process 2
ship: R/V Roger A. Revelle
Methodology:
Samples that had been frozen in liquid nitrogen were extracted as described
previously (Goericke and Repeta, 1993, MEPS 101: 307 - 313).
Pigment extracts were analyzed on the reverse phase HPLC system described
below. Pigments were detected by absorbance at 440nm. The identity of
pigments was confirmed by frequently checking on-line spectra against those
of standards.
Column - Alltech Adsorbosphere C18, HS 3um, 4.6 mm, 10 cm.
Solvents - A [methanol : acetonitrile : water : 0.5M aq ammonium acetate
(30:30:30:10)
B [methanol : acetonitrile : ethyl acetate (10:35:55)].
Gradient - [time (min); solvent A, solvent B] - [0; 100, 0], [4; 75, 25],
[24; 12, 88], [26; 12, 88], [28; 100, 0].
Instruments - Waters 510 pumps, Shimadzu autosampler and uv/vis detector,
Waters 991M photodiode detector and Waters Millenium data system