Hydrocasts: CTD and Rosette Sample
During each cruise, a minimum of four hydrocasts were performed to collect a suite of core monthly observations. Additional hydrocasts were performed for specific process studies. We conducted separate shallow and deep casts to obtain better vertical resolution for chemical observations, and for productivity and pigment observations. Water was collected with a SeaBird rosette equipped with 12 (8 liter) teflon-coated Niskin bottles (bottle springs were also teflon-coated) at 20 depths between the surface and 1310 m. The rosette housed the CTD, which collected continuous profiles of temperature and salinity. The CTD also had a SBE-43 oxygen probe, a Wetlabs ECO fluorometer outfitted for chlorophyll-a estimates, and a C-Star transmissometer (660 nm, Wetlabs). Beam attenuation measurements were added to the time series on its 11th cruise (November 1986) originally using a SeaTech transmissometer. The rosette was controlled with a SeaBird deck unit via conducting cable, but alternatively it had been actuated automatically based on pressure recordings via an Autofire Module (SBE AFM) when breaks in cable conductivity had occurred.
Between November 1995 and September 1996, three separate SBE-19 CTDs were used in repeated casts until a reliable salinity profile was obtained below the oxycline. The SBE-19 model CTDs frequently failed to provide reliable conductivity values below the oxycline in the Cariaco Basin. Starting in September 1996, the SBE-19 CTDs were replaced by SBE-25 CTDs, which provided extremely accurate and reliable data in anoxic waters.
All CTDs were calibrated at the Sea-Bird factory once per year. The accuracy of the pressure sensor was 3.5 m and had a resolution of 0.7 m. The temperatures accuracy was 0.002°C with a resolution of 0.0003°C. The conductivity accuracy was 0.003 mmho/cm with a resolution of 0.0004 mmho/cm.
Discrete Salinity
Continuous salinity profiles were calculated from the CTD measurements. Discrete salinity samples were analyzed using a Guildline Portasal 8410 salinometer standardized with IAPSO Standard Seawater, with a precision of better than ± 0.003 and a resolution of 0.0003 mS/cm at 15° C and 35 psu, the accuracy was ±0.003 at the same set point temperature as standardization and within -2° and +4°C of ambient. These salinity values were used to check, and when necessary calibrate, the CTD salinity profiles.
Discrete Oxygen
Continuous dissolved oxygen (O2) profiles were obtained with a SBE-43 Dissolved Oxygen Sensor coupled to the SBE-25 CTD. Discrete oxygen samples were collected in duplicate using glass-stoppered bottles and analyzed by Winkler titration (Strickland and Parsons, 1972, as modified by Aminot, 1983). The analytical precision for discrete oxygen analysis was ±3 mM, based on analysis of duplicate samples, with a detection limit of 5 mM. The in-situ oxygen values were used to check, and when necessary calibrate, the CTD oxygen profiles.
Discrete Chlorophyll
Continuous fluorescence profiles were obtained with a Sea Tech Fluorometer coupled to the SBE-25 CTD. Discrete chlorophyll samples were immediately filtered through 25 mm Whatman GF/F filters in triplicate and frozen. The chlorophyll-a determination followed Holm-Hansen et al. (1965), and the calculations were done as indicated in Lorenzen (1966). Among the modifications are the use of methanol instead of acetone as an extraction solvent due to its greater efficiency (Holm-Hansen and Riemann, 1978) and the use of a sonic dismembrator (Wright et al., 1997). The method applies to all ranges of chlorophyll-a concentration found in seawater. The method detection limit is 0.01 μg L-1 for natural waters (for a 0.5 L sample). Fluorescence profiles were calibrated with in-situ chlorophyll-a and it is presented in its native units of RFU (fluor_CTD) and also in micrograms/m^3 (fluor_chla).