The data were collected using the Wecoma's SBE911+ system with dual Temperature and Conductivity
sensors, a SeaTech fluorometer sensor a Biospherical Instruments/LICOR PAR (Irradiance) sensor,
and a Sea Bird SBE043 dissolved oxygen sensor.
Link to NODC Data Form 24-13 for Wecoma cruise number W0308C CTD Data Submission
File: w0308c_con_rept.txt
Date: 11/30/2008
Instrument configuration file: F:confilesw0308ccast003.CON
Configuration report for SBE 911plus/917plus CTD
------------------------------------------------
Frequency channels suppressed : 0
Voltage words suppressed : 0
Computer interface : IEEE-488 (GPIB)
Scans to average : 1
Surface PAR voltage added : No
NMEA position data added : Yes
Scan time added : No
1) Frequency 0, Temperature
Serial number : 1366
Calibrated on : 05-Sept-02
G : 4.83426629e-003
H : 6.76281636e-004
I : 2.56989753e-005
J : 1.99492586e-006
F0 : 1000.000
Slope : 1.00000000
Offset : 0.0000
2) Frequency 1, Conductivity
Serial number : 2356
Calibrated on : 17-Dec-02
G : -1.00784754e+001
H : 1.46955327e+000
I : 4.36564460e-004
J : 6.08868326e-005
CTcor : 3.2500e-006
CPcor : -9.57000000e-008
Slope : 1.00000000
Offset : 0.00000
3) Frequency 2, Pressure, Digiquartz with TC
Serial number : 9P6012-2843
Calibrated on : 17-Jul-01
C1 : -4.029491e+004
C2 : -8.058990e-001
C3 : 1.260260e-002
D1 : 3.076300e-002
D2 : 0.000000e+000
T1 : 3.021184e+001
T2 : -4.772503e-004
T3 : 3.981640e-006
T4 : 3.194480e-009
T5 : 0.000000e+000
Slope : 1.00000000
Offset : 0.00000
AD590M : 1.157000e-002
AD590B : -8.237500e+000
4) Frequency 3, Temperature, 2
Serial number : 1371
Calibrated on : 17-Dec-02
G : 4.83369723e-003
H : 6.78344390e-004
I : 2.65701597e-005
J : 2.09190892e-006
F0 : 1000.000
Slope : 1.00000000
Offset : 0.0000
5) Frequency 4, Conductivity, 2
Serial number : 1054
Calibrated on : 02-Jul-02
G : -4.08915197e+000
H : 5.91598520e-001
I : -8.13084544e-005
J : 3.90272813e-005
CTcor : 3.2500e-006
CPcor : -9.57000000e-008
Slope : 1.00000000
Offset : 0.00000
6) A/D voltage 0, Fluorometer, Seatech/Wetlabs FLF
Serial number : 101S
Calibrated on : 19 Dec 01
Scale factor : 3.000000e+000
Offset : 0.000
7) A/D voltage 1, Transmissometer, Chelsea/Seatech/Wetlab CStar
Serial number : CTS-590DR
Calibrated on : 05/01/02
M : 1.0000
B : 0.0000
Path length : 0.250
8) A/D voltage 2, Oxygen, SBE 43
Serial number : 0387
Calibrated on : 03/01/03
Equation : Owens-Millard
Coefficients for Owens-Millard:
Soc : 4.3780e-001
Boc : 0.0000
Offset : -0.4728
Tcor : 0.0015
Pcor : 1.35e-004
Tau : 0.0
Coefficients for Murphy-Larson:
Soc : 0.00000e+000
Offset : 0.00000e+000
A : 0.00000e+000
B : 0.00000e+000
C : 0.00000e+000
E : 0.00000e+000
Tau : 0.00000e+000
9) A/D voltage 3, Free
10) A/D voltage 4, Altimeter
Serial number : Datasonics
Calibrated on :
Scale factor : 15.000
Offset : 0.000
11) A/D voltage 5, Free
12) A/D voltage 6, PAR/Irradiance, Biospherical/Licor
Serial number : 4246
Calibrated on : 25Jul2000
M : 1.00000000
B : 0.00000000
Calibration constant : 203000000000000.00000000
Multiplier : 53.95400000
Offset : 0.00000000
13) A/D voltage 7, Free