Dataset: cal_iso
Data Citation:
Granger, J., Bourbonnais, A., Wilson, S. (2019) Calibration of the Isotope Ratio Mass Spectrometer vs. aliquots of 15N2 gas diluted in air then helium. Biological and Chemical Oceanography Data Management Office (BCO-DMO). (Version 1) Version Date 2019-10-02 [if applicable, indicate subset used]. doi:10.1575/1912/bco-dmo.778000.1 [access date]
Terms of Use
This dataset is licensed under Creative Commons Attribution 4.0.
If you wish to use this dataset, it is highly recommended that you contact the original principal investigators (PI). Should the relevant PI be unavailable, please contact BCO-DMO (info@bco-dmo.org) for additional guidance. For general guidance please see the BCO-DMO Terms of Use document.
DOI:10.1575/1912/bco-dmo.778000.1
View, Subset and Download Data
BCO-DMO is preparing to move to a new data access system that provides more functionality and features. This system, called ERDDAP, is free and open source with a strong and growing user community. Currently, we are providing the following data access and download capabilities for datasets as we work through our legacy data holdings. These capabilities will soon be available for all data at BCO-DMO:Data Access
- - view the data in an HTML table
- - filter the data before viewing or downloading in a variety of formats
- - download the data with comma-separated values (Excel-ready)
- - download the data with tab-separated values (Excel-ready)
- - download the data as GeoJSON. Try it out at geojson.io
- - download the data as a valid MATLAB file
- - download the data in NetCDF format
- - download the data in Ocean Data View format
Give Us Feedback
Do you have thoughts, questions or constructive feedback about data access at BCO-DMO? Let us know: feedback [at] bco-dmo [dot] org
Principal Investigator:
Julie Granger (University of Connecticut, UConn)
Co-Principal Investigator:
Annie Bourbonnais (University of Massachusetts Dartmouth, UMass Dartmouth)
Samuel Wilson (University of Hawai'i, UH)
BCO-DMO Data Manager:
Mathew Biddle (Woods Hole Oceanographic Institution, WHOI BCO-DMO)
Version:
1
Version Date:
2019-10-02
Restricted:
No
Validated:
Yes
Current State:
Final no updates expected
Calibration of the Isotope Ratio Mass Spectrometer vs. aliquots of 15N2 gas diluted in air then helium
Abstract:
In order to verify that measurements of the 15N2 atom% of dissolved N2 gas samples effectuated by Membrane Inlet Mass Spectrometry are accurate, we compared these to corresponding measurements of parallel samples on a Delta V Isotope Ratio Mass Spectrometer (IRMS). To this end, we first ensured that the IRMS returns accurate 15N2 atom% values, by measuring the 15N2 atom% of a serial dilutions of a 15N2 gas stock in air and helium. This calibration is documented herein, showing that IRMS measurements in the experimental range were accurate, as expected